[Reader Insight] Selection of Chromatography Detectors Based on Molecular Structure in Analytical Method Development

  • Home
  • >
  • Blogs
  • >
  • [Reader Insight] Selection of Chromatography Detectors Based on Molecular Structure in Analytical Method Development

This article is written by an expert chromatographer under the pen name of Chromatography Mound. Welch Materials, Inc. is authorized to translate this article to English and publish it on behalf of the author.

Introduction

When tasked with analyzing a compound or a group of compounds within a specific matrix, how do we design a preliminary and comprehensive experimental protocol?

To answer this question, we first recall that a complete analytical experiment involves four essential steps: sample preparation, pretreatment, separation, and detection. Each of these steps includes numerous intricate details.

In this article, we will focus on one critical aspect: how to select the appropriate detector based on the molecular structure of the analyte.

Detector Selection Based on Functional Groups

The core of analytical method development lies in the selection of tools, such as pretreatment techniques and detectors, that demonstrate high selectivity for the target analyte. Starting from the molecular structure enables us to identify the most suitable tools. 

Selectivity is always relative; here, it refers specifically to the target compound.

The first step in developing an analytical method is to generate a chromatogram of a standard solution. In this step, the detector is determined based on the functional groups present in the molecule since the detection mechanism is closely related to these groups.

Compounds with Conjugated Structures:
If the molecule contains extensive conjugated structures—such as double bonds, triple bonds, or benzene rings—diode-array detectors (DAD) or fluorescence detectors (FLD) are suitable options. The stronger the conjugation, the greater the detector's response.

Compounds Without Conjugated Structures:
For molecules lacking conjugated structures, detectors such as differential refractive index (RID), evaporative light scattering (ELSD), or gas chromatography flame ionization detectors (GC-FID) may be more appropriate.

Compounds Containing Halogens:
For molecules with halogens like fluorine (F), chlorine (Cl), or bromine (Br), gas chromatography electron capture detectors (GC-ECD) are effective choices.

Trace-Level Compounds:
When the analyte concentration is very low, mass spectrometry (MS) is preferred. For LC-MS using electrospray ionization (ESI), the molecule must contain nucleophilic or electrophilic groups such as amines, carbonyls, carboxyls, or sulfonates to achieve a good response. Otherwise, GC-MS is perferred, yet the compound should have a low boiling point, typically below 350 °C.

Case Study: Detector Selection for Specific Compounds

  • Benzoic Acid and Sorbic Acid:
    Both compounds contain conjugated structures such as benzene rings or double bonds, as well as electrophilic groups like carboxyl groups. They can be analyzed using LC-PDA or LC-MS. However, since these compounds are often stabilized as salts in samples, when analyzing using GC methods, pH adjustment of the extraction solution to acidic conditions is necessary to ensure their presence in molecular form and lower their boiling points.
  • TBHQ and Ethylparaben:
    The structures of tert-butylhydroquinone (TBHQ) and ethylparaben both feature conjugated benzene rings. Their boiling points are 291.4 °C and 297.5 °C, respectively, and they usually exist in molecular form. These compounds can be analyzed using LC-PDA or GC-FID.
  • Polychlorinated Biphenyls (PCBs):
    PCBs are a group of compounds with multiple isomers, consisting of two benzene rings and several chlorine atoms. Even for highly chlorinated PCBs, the boiling points are very low (e.g., hexachlorobiphenyls have their boiling points ranging from 29 to 33 °C). Common PCBs can be analyzed using GC-ECD. LC-PDA, FLD, or GC-MS are also viable options.

Considerations Beyond Molecular Structure

Establishing a standard chromatogram is only the first step in developing an analytical method. When selecting a detector, apart from molecular structure, it is also essential to consider the sample's background matrix, which is strongly related to factors like the sample quantity and dilution ratio. For common instruments, LC-PDA, RI, ELSD, and GC-FID detectors typically require analyte concentrations to be in ppm level, while GC-MS, GC-ECD, and LC-FLD detectors can achieve detection limits in ppb level. 

The distinction on analyte concentration is as crucial as that on molecular structure when choosing the appropriate detector for a specific analytical method.